FLOWING GAS CORROSION TEST CABINET



특징
- test atmosphere: 1±0.01(m /h) or 3, 1.5±0.03(m /h) 3
- Class-M : 0.5ppm to 25ppm or more
- Class-L : 50ppb to 2500ppb (2.5ppm)
- Class-VL : 5ppb to 250ppb
SAFETY & STABILITY
for long time unmanned running by using
- Air suction and cleaning system by CONSTANT-FLOW WASTE GAS WASHING PUMP: Model GA-101(W)
- Accurate and stable micro flow rate of test gas supplying under atmospheric pressure by YAMASAKI CONSTANT-FLOW GAS PUMP: Model VCP-2MF (0.01 to 2.50L/h)/±0.5%
- MAGNETIC GAS REDUCING VALVE: Model WR-13S
MEETS TO WORLDWIDE MFG CORROSION TEST STANDARDS
for Electric-electronics products
Class-M
IEC 60068-2-42, Test Kc 25±5ppmSO , 25±2℃, 75(70 to 80)%RH, 21da 2 ys
IEC 60068-2-43, Test Kd 10 to 15ppmH S, 25±2℃, 75(70 to 80)%RH, 21days
DIN 40046-36 10±2ppmSO , 25±2℃, 75±5%RH, 21days
DIN 40046-37 1±0.3ppmH S, 25±2℃, 75±5%RH, 21days
JEIDA 25 (Japan) 3±1ppmH S, 40±2℃, 85±5%RH
JEIDA 32 (Japan) 10±3ppmSO , 40±2℃, 85±5%RH
(Japan) 3ppmH S+10ppmSO ,40℃,85%RH
Class-L
ISO 10062-Method A 0.5ppmSO , 25℃, 75%RH, max.90days
ISO 10062-Method B 0.1ppmH S, 25℃, 75%RH, max.90days
ISO 10062-Method C
IEC 60068-2-60/Ke method-1
JIS C 0048(1)
0.5ppmSO +0.1ppmH S, 25℃, 75%RH
TECHNICAL DETAILS: Model GH-180
DIMENSIONS
TEST CABINET
FLOOR AREAS
500(W) x 500(D) x 720(H)mm
MAIN BODY 1210(W) x 900(D)mm
GA-101 500(W) x 600(D)mm
TOTAL HEIGHT
TOTAL NET MASS
POWER REQUIREMENT
TEST GAS CONCENTRATIONS
approx. 2300mm
approx. 380kg
2.5kw single phase (100V AC), 50 or 60Hz
Class-M(by 1%), Class-L(by 0.1%), or Class-VL(by 0.01%) standard gas
FLOWING TEST ATMOSPHERE 1000±10(L/h)…approx. 5.6 changes/h or
1500±30(L/h)…approx. 7.5 changes/h
TEST TEMPERATURE
TEST HUMIDITY
AIR CONDITIONER
MATERIALS
+20℃ to 50℃ (±0.5℃)
75% RH to 95% RH (±3% RH)
0.7kw air-cooled refrigerator
TEST CABINET - Chlorinated polyvinyl chloride (CPVC)
DOOR - SEAL Viton rubber
TEMP. PROBE - Pyrex glass/Pt-resistance
GAS TUBING - PTFE(Teflon) and Tygon tube
AMBIENT CONDITION 18 to 28℃ under running (recommended)
TYPICAL ACCESSORIES MAGNETIC GAS REDUCING VALVE (Model WR-13S)
CONSTANT-FLOW GAS MICRO-PUMP (Model VCP-2MF)
CONSTANT-FLOW WASTE GAS WASHING PUMP (Model GA-101)

FLOW DIAGRAM FOR A FLOWING MIXED GAS (MFG) TEST