세상을 향한 발돋움으로 글로벌 리더 ABCMTS㈜가 되겠습니다.
Class-M | |
---|---|
IEC 60068-2-42, Test Kc | 25±5ppmSO , 25±2℃, 75(70 to 80)%RH, 21da 2 ys |
IEC 60068-2-43, Test Kd | 10 to 15ppmH S, 25±2℃, 75(70 to 80)%RH, 21days |
DIN 40046-36 | 10±2ppmSO , 25±2℃, 75±5%RH, 21days |
DIN 40046-37 | 1±0.3ppmH S, 25±2℃, 75±5%RH, 21days |
JEIDA 25 (Japan) | 3±1ppmH S, 40±2℃, 85±5%RH |
JEIDA 32 (Japan) | 10±3ppmSO , 40±2℃, 85±5%RH |
(Japan) | 3ppmH S+10ppmSO ,40℃,85%RH |
Class-L | |
ISO 10062-Method A | 0.5ppmSO , 25℃, 75%RH, max.90days |
ISO 10062-Method B | 0.1ppmH S, 25℃, 75%RH, max.90days |
ISO 10062-Method C IEC 60068-2-60/Ke method-1 JIS C 0048(1) |
0.5ppmSO +0.1ppmH S, 25℃, 75%RH |
DIMENSIONS TEST CABINET FLOOR AREAS |
500(W) x 500(D) x 720(H)mm MAIN BODY 1210(W) x 900(D)mm GA-101 500(W) x 600(D)mm |
---|---|
TOTAL HEIGHT TOTAL NET MASS POWER REQUIREMENT TEST GAS CONCENTRATIONS |
approx. 2300mm approx. 380kg 2.5kw single phase (100V AC), 50 or 60Hz Class-M(by 1%), Class-L(by 0.1%), or Class-VL(by 0.01%) standard gas |
FLOWING TEST ATMOSPHERE | 1000±10(L/h)…approx. 5.6 changes/h or 1500±30(L/h)…approx. 7.5 changes/h |
TEST TEMPERATURE TEST HUMIDITY AIR CONDITIONER MATERIALS |
+20℃ to 50℃ (±0.5℃) 75% RH to 95% RH (±3% RH) 0.7kw air-cooled refrigerator TEST CABINET - Chlorinated polyvinyl chloride (CPVC) DOOR - SEAL Viton rubber TEMP. PROBE - Pyrex glass/Pt-resistance GAS TUBING - PTFE(Teflon) and Tygon tube |
AMBIENT CONDITION | 18 to 28℃ under running (recommended) |
TYPICAL ACCESSORIES | MAGNETIC GAS REDUCING VALVE (Model WR-13S) CONSTANT-FLOW GAS MICRO-PUMP (Model VCP-2MF) CONSTANT-FLOW WASTE GAS WASHING PUMP (Model GA-101) |